Cardiovascular
Phlebology
Set for venous insufficiency treatment, FLEBOGRIF
Medical use
Ablation of lower limb varicose veins
Ablation of lower limb varicose veins
The set for venous insufficiency treatment, Flebogrif is intended for mechanochemical ablation of lower extremity superficial truncal veins.
The set Flebogrif consists of:
- catheter;
- needle;
- guidewire;
- introducer sheath;
- dilator;
- three-way stopcock;
- syringes.
The catheter consists of two coaxially cooperating parts:
- internal - designed to be a lumen for the guidewire and to deliver the foamed sclerosant;
- external - with length markers every 1 cm (starting at 10 cm from the distal end).
The catheter is equipped with a cutting elements consisting of five retractable, branching arms arranged circumferentially designed to scratch the inner wall of the vein.
Intended purpose:
The Set for Venous Insufficiency treatment, Flebogrif is intended for mechanochemical ablation of lower extremity superficial truncal veins.
-
Download patient
brochure
Technical data overview
Catheter diameter | 6 F (2.08 mm) |
---|---|
Catheter length | 60 cm or 90 cm |
Catheter dead space volume | 0.8 ml for 60 cm catheter; 1 ml for 90 cm catheter |
Needle size | 18 G |
Needle length | 7 cm |
Guidewire diameter | 0.035’’ |
Guidewire length | 150 cm for 60 cm catheter; 180 cm for 90 cm catheter |
Guidewire coating | PTFE (polutetrafluoroethylene) |
Introducer sheath with dilator diameter | 6 F (minimum ID 2.18 mm) |
Syringe volume | 5 ml |
Syringe connector type | luer-lock |
Available versions
Catalogue No. | Catheter size [F] |
Catheter length [cm] |
---|---|---|
FLE6F60 | 6 | 90 |
FLE6F90 | 6 | 60 |
Set elements
-
Components of the Set for venous insufficiency treatment Flebogrif
- Flebogrif catheter 6 F
- Needle 18 G
- J-type guidewire 0.035”
- Introducer sheath 6 F
- Dilator
- Three-way stopcock
- Luer-lock syringe 5 ml 2 pcs
-
Flebogrif catheter
- Cutting elements
- Inner shaft
- Outer sheath
- Soft tip
- Length markers
- Size information
- Clamping knob (Y connector)
- Inner shaft hub with the dead space information
- Sclerosant injection port (female luer-lock connector)
- Flushing port (female luer-lock connector)
L – Effective (usable) length